Page 27 - Workswell Catalogue
P. 27

MEDICAS Main features


          THE MOST IMPORTANT FUNCTIONS & PARAMETERS                                                                26





            ADJUSTABLE LEVELS - FEVER, ELEVATED AND NORMAL TEMPERATURE
            For  clear  visualization,  Workswell  MEDICAS  works   This  way  of  displaying  allows  to  optically  filter  out
            with three levels:                                situations where there is no elevated temperature in
            1. Gray-white palette is used for normal body tem-  the image and, on the contrary, to clearly and clear-
              peratures (and ambient temperatures).           ly identify the situation when a person with elevated
            2. Yellow range is used for elevated temperature.  temperature or fever occurs.
            3. Red is for fever signaling.
              The  adjustable  levels
              settings  are  shown  on
              the temperature scale.



          FEVER SCREENING STANDARDS: IEC 80601-2-59 AND ISO/TR 13154
          World’s two top international standards groups, the IEC and ISO, have published standards covering fever (i.e., fe-
          brile) screening. While IEC 80601-2-59 is focused on essential performance of screening cameras, ISO/TR 13154:2017
          is operational guidelines for identifying febrile humans using a screening camera.
          The  Workswell  MEDICAS  thermal  im-
          ager together with the Worksell BB200
          black  body  have  been  designed  to
          meet the requirements of these stan-
          dards as much as possible. That’s why
          Workswell  MEDICAS  offer  parame-
          ters:  uncertainty  ±  0.3  °C  (±  0.54  °F),                                                             Medical
          sensitivity  0.03  mK,  high  resolution
          640  x  512  px,  but  also  features  like:
          threshold  temperature  alarm  system,
          start-up warning and more.



          WHY WORKSWELL MEDICAS CAMERA                        HUMAN TEMPERATURE SCREENING
          Standard infrared LWIR measurement works on ± 2 °C   How the different resolution influences the final tem-
          (± 3.6 °F) accuracy, which is absolutely not acceptable   perature value
          for this application.
          Workswell MEDICAS is a unique infrared system which
          is calibrated in real time on a black body which pro-
          vides outstanding accuracy ± 0.3 °C (± 0.54 °F) and high
          infrared sensor resolution 640 x 512 px.

          Thanks to this precision in measurement the fever could
          be detected by providing double visual alarm on:
             † warning temperature
             † critical temperature
             † activate digital triggers and alarms
          An operator of the system can easily recognize a person
          with the fever.