MEDICAS Main features
THE MOST IMPORTANT FUNCTIONS & PARAMETERS 26
ADJUSTABLE LEVELS - FEVER, ELEVATED AND NORMAL TEMPERATURE
For clear visualization, Workswell MEDICAS works This way of displaying allows to optically filter out
with three levels: situations where there is no elevated temperature in
1. Gray-white palette is used for normal body tem- the image and, on the contrary, to clearly and clear-
peratures (and ambient temperatures). ly identify the situation when a person with elevated
2. Yellow range is used for elevated temperature. temperature or fever occurs.
3. Red is for fever signaling.
The adjustable levels
settings are shown on
the temperature scale.
FEVER SCREENING STANDARDS: IEC 80601-2-59 AND ISO/TR 13154
World’s two top international standards groups, the IEC and ISO, have published standards covering fever (i.e., fe-
brile) screening. While IEC 80601-2-59 is focused on essential performance of screening cameras, ISO/TR 13154:2017
is operational guidelines for identifying febrile humans using a screening camera.
The Workswell MEDICAS thermal im-
ager together with the Worksell BB200
black body have been designed to
meet the requirements of these stan-
dards as much as possible. That’s why
Workswell MEDICAS offer parame-
ters: uncertainty ± 0.3 °C (± 0.54 °F), Medical
sensitivity 0.03 mK, high resolution
640 x 512 px, but also features like:
threshold temperature alarm system,
start-up warning and more.
WHY WORKSWELL MEDICAS CAMERA HUMAN TEMPERATURE SCREENING
Standard infrared LWIR measurement works on ± 2 °C How the different resolution influences the final tem-
(± 3.6 °F) accuracy, which is absolutely not acceptable perature value
for this application.
Workswell MEDICAS is a unique infrared system which
is calibrated in real time on a black body which pro-
vides outstanding accuracy ± 0.3 °C (± 0.54 °F) and high
infrared sensor resolution 640 x 512 px.
Thanks to this precision in measurement the fever could
be detected by providing double visual alarm on:
warning temperature
critical temperature
activate digital triggers and alarms
An operator of the system can easily recognize a person
with the fever.