Black body BB200

    27
          HIGH ACCURATE AREA BLACK BODY RADIATION SOURCE




                                                               BB200 Key features description
                                                               Continuous temperature   30 °C – 200 °C (86 °F to 392 °F) with 0.1 °C (°F) single step
                                                               settings of BB200
                                                                               100 mm stabilized diameter area with pyramid shapes for
                                                               Big measurement area  increasing stability of emission and decreasing the reflective
                                                                               effects
                                                                               Temperature stability of the central area better than:
                                                                               • ± 0.05 °C / hour at 40 °C set-value (± 0.09 °F / hour at
                                                                                104 °F set-value)
                                                               Outstanding stability  • ± 0.06 °C / hour at 100 °C set-value( ± 0.11 °F / hour at
                                                                                212 °F set-value)
                                                                               • ± 0.07 °C / hour at 150 °C set-value (± 0.13 °F / hour at
                                                                                302 °F set-value)
                                                               BB200 technical specification
                                                               Surface emissivity  0.95 ± 0.005 for LWIR spectrum
                                                                               30 °C to 200 °C in continuous steps by 0.1 °C (86 °F to 392 °F
                                                               Temperature range  by 0.1 °F)
                                                                               (the lowest set-value must be higher about 5 °C (9 °F) from
                                                                               ambient temperature)
                                                               Aperture diameter  100 mm
                                                                               Better than ± 0.08% (Precise PID regulation with 16 single
          Workswell BB200 is a high accurate area black body ra-  Long-term time stability  thermal elements, measured/checked by certified external
          diation source for infrared thermography. BB200 offer-  Space thermal non-uni-  PT100 probe)
                                                                               ≤ 0,3 °C (0.5 °F) (temperature dispersion)
          ing continuous temperature settings from 30 °C - 200 °C   formity    USB-C, RS485 control serial interface cable (optional)
                                                               Control interface
          (86 °F - 392 °F) with 0.1 °C (0.18 °F) single step. Surface   Display  OLED Display: real-time temperature measurement, set-val-
          is 100 mm diameter temperature stabilized area with                  ue, visualization
                                                                               Set temperature value can be saved to the memory and au-
          pyramid shapes for increasing stability of emission and   Temperature memory  tomatically loaded when BB200 connected to power supply,
                                                                               memory function available
          decreasing  the  reflective  effects.  Surface  emisivity  is   LED signalization  LED signalization for different modes (warming, stabilization,
          0.95 ± 0.005 for LWIR spectrum. Long-term time stabili-  Temperature units  cooling, electronics overheating)
                                                                               Celsius and Fahrenheit (user configurable)
          ty is better than ±0.08 % (precise PID regulation with 16   Power supply, weight & dimensions
          single thermal elements,measured/checked by certified   Input supply voltage  24 VDC, Coaxial 2 x 6.4 mm, outer shell – GND, power supply
                                                                               adapter 110/230 VAC included in the package
          external PT100 probe).                               Power consumption (avg.)  24 W (up to 100 °C temp. range), 96 W (up to 200 °C temp.
                                                                               range)
          Temperature value can be saved to the memory and     Weight          < 1 200 grams without the tripod (tripod holder included in
                                                                               the package)
          automatically  loaded  when  the  black  body  BB200  is   Dimensions (Diameter,   117 mm x 88 mm (without aluminium stand or cabling)
                                                               Depth)
          connected  to  power  supply.  Memory  function  is  also   Mounting holes  1 x 1/4-20 UNC thread
    Medical
          available. Mouting hole is a very standard 1/4-20 UNC   Housing material  Special Teflon body with aluminum parts
                                                               Environmental
          thread. Housing material is a special teflon body with   Operating temperature   10 °C to +30 °C (50 °F to +86 °F)
          aluminum  parts.  The  BB200  is  also  equipped  with  a   range    -30 °C to +60 °C (-22 °F to +140 °F)
                                                               Storage temperature range
          OLED display for visualizing the set and actual surface   Relative humidity  Non condensing conditions
          temperature values.                                  In the package   Black body, calibration certificate, ball adapter, power
                                                                               supply, transport case
          This blackbody meets the demand for a simple, cost effective but high accuracy calibrator for the calibration of
          thermal cameras used to screening febrile subjects. This is also the reason why this black body is used with the
          Workswell MEDICAS thermal imaging system and is part of the standard packaging.






















                    Black body temperature stability                        Black body surface detail
   23   24   25   26   27   28   29   30   31   32   33